ANALYSIS OF STRENGTH CHARACTERISTICS OF CEMENT-SCREW FIXATION OF MULTI-UNIT ABUTMENTS TO DENTAL IMPLANTS WITH FLAT AND CONICAL TYPE OF CONNECTION IN EXPERIMENT

https://doi.org/10.51922/1818-426X.2025.4.118

Abstract

Purpose is to analyze the strength characteristics of the cement-screw fixation of multiunit abutments to dental implants with a planar and conical type of connection in an experiment under conditions of cyclic loads simulating functional loads in the oral cavity.
Methods. The strength characteristics of the cement-screw connection (CSC) of multiunit abutments to dental implants made of Ti-6Al-4V alloy with a diameter of 3.75 mm and a length of 11.5 mm with a conical and planar type of connection were evaluated. A total of 48 dental implants were included in the experiment – 24 with a planar and 24 with a conical type of connection. In turn, every 24 implants were divided into six groups depending on the experimental conditions: three groups with a loading angle of 22.5° and a force of 100 N with a number of loading cycles of 100 000, 150 000 and 200 000; three groups at loading angle of 45° and impact force of 100 N with number of loading cycles of 100 000, 150 000 and 200 000.
Results. In comparative analysis of the strength of experimental samples under maximum load conditions (loading angle of 45° and 200,000 loading cycles) in the 2nd position of the implant, as the weakest element of the entire structure, it was revealed that with the conical type, the depth of the CSC decreases by 19 %, and the degree of screw clamping – by 73 % from the initial level to the load, while with the planar type, similar parameters decrease by 33 % and 80 %, respectively.
Conclusion. It has been established that the conical type of connection is more reliable than the planar type, its use allows you to increase the resistance to loads, since the accumulation of fatigue of orthopedic structures with a conical multi-unit connection increases in 1.2 times, with a planar multi-unit connection – in 1.5 times. compound simulating the projection of the 2nd premolar) by 24 % compared to the planar type of compound.

About the Authors

S. V. Pryalkin
Educational Institution «Belarusian State Medical University»
Belarus


S. P. Rubnikovich
Educational Institution «Belarusian State Medical University»
Belarus


V. N. Busco
Institute of Applied Physics of the National Academy of Sciences of Belarus
Belarus


References

1. Pakhshan, G., Zahra B. Multi-Unit Abutment: Why, Where, When? // J. Oral Med. and Dent. Res. – 2025. – Vol. 6, № 2. – P. 1–8.

2. Andriyanova, N. V. Uluchshenie konstrukcii vnutrikostnogo dental’nogo vintovogo implantata // Tendencii razvitiya nauki i obrazovaniya. – 2017. – № 33–1. – S. 31–32.

3. Farah, R. I., Alshabi A. M. Technique to verify the accuracy of a definitive cast before the fabrication of a fixed dental prosthesis // J. Prosthet. Dent. – 2016. – Vol. 116, № 3. – Р. 325–327.

4. Baig, M. Multi-unit implant impression accuracy: A review of the literature // Quintessence international. – 2014. – Vol. 45. – P. 39–51. – doi: org/10.3290/j.qi.a 30769.

5. Wu, Y.-L., Aaron Y.-J. W. A method of fabricating an accurate repositioning device for relocating multiple multiunit abutments // The Journal of Prosthetic Dentistry. – 2017. – Vol. 118, № 4. – Р. 564–566.

6. Albadani, M., Alahmari M. Full-Mouth Rehabilitation Using Combined Multiunit Abutment and Screw-Retained T-Base Abutment System: A Case Report with 24-Month Followup // Open Dent. J. – 2024. – Vol. 18. – P. e18742106293859. – doi: org/10.2174/0118742106293859240520103643.

7. Mashkov, K. N. Primenenie dental’nyh implantatov pri polnom otsutstvii zubov // Byulleten’ medicinskih internet-konferencij. – 2019. – T. 9, № 6. – S. 243–244.

8. Bus’ko, V. N. Laboratornaya ustanovka dlya issledovaniya ustalostnoj povrezhdaemosti ploskih ferromagnitnyh obrazcov // Pribory i tekhnika eksperimenta. – 2011. – № 1. – S. 165–167


Review

For citations:


Pryalkin S.V., Rubnikovich S.P., Busco V.N. ANALYSIS OF STRENGTH CHARACTERISTICS OF CEMENT-SCREW FIXATION OF MULTI-UNIT ABUTMENTS TO DENTAL IMPLANTS WITH FLAT AND CONICAL TYPE OF CONNECTION IN EXPERIMENT. Medical Journal. 2025;(4):118-124. (In Russ.) https://doi.org/10.51922/1818-426X.2025.4.118

Views: 55

JATS XML


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 1818-426X (Print)